Storage
Reference: storage temperatures for peptide reagents
12 November 2026 · 5 min read
Temperature is the storage variable a laboratory controls continuously, and the one most often decided by whichever shelf happened to be free. This note sets out the usual zones, what each is used for, and the two details that cause more trouble than the zone itself: the form of the material, and the date the vial was first opened.
The zones
| Zone | Nominal range | Normally used for |
|---|---|---|
| Ambient | 15 to 25 °C | sealed lyophilised solids in transit, unopened solvent bottles, consumables |
| Refrigerated | 2 to 8 °C | an opened multiple entry solvent, a working stock in active use, a solution held between two sessions in the same week |
| Freezer | -20 °C | longer holding of sealed lyophilised solids, and of solutions already divided into portions |
| Deep freezer | -80 °C | long holding of aliquoted solutions, in facilities that have one |
Read the table as a map of what each zone is for, not as a shelf life table. How long a given material tolerates a given zone belongs to the documentation supplied with it. The quality page sets out what accompanies a catalogue item.
Two points about hardware rather than numbers. A nominal minus 20 degrees Celsius describes the air in the cabinet, not a vial pressed against the door seal. And a zone is only as good as its monitoring: a cabinet with no thermometer and no log records its failures nowhere, so a power cut over a weekend surfaces a month later as unexplained results.
A sealed solid and a solution are two different materials
Most of what degrades a peptide needs water to do it. Lyophilisation removes that water, which is the point of supplying the material as a solid. A sealed solid, kept dry under the headspace it was packed with, is comparatively insensitive to a short excursion in temperature.
Reconstitution reverses that. The moment the solvent goes in, hydrolysis, oxidation, adsorption onto the container wall and, in a non preserved solvent, microbial growth all become live concerns. The same substance now needs a colder zone, a shorter horizon and division into portions, so that it is thawed once rather than ten times.
The practical rule that follows: the form of the material, not its name on the label, decides where it is stored. Solvent choices are set out under bacteriostatic water and sterile water in the catalogue.
Why a freezer that defrosts itself is a poor choice
A frost free domestic freezer keeps itself clear of ice by warming its evaporator, and usually part of the cabinet, on a repeating cycle through the day. That is a feature for food and a defect for reagents: every cycle is a small warming phase followed by a return to temperature, which is exactly the thermal cycling that cold storage exists to avoid.
For a solution, each cycle is a partial thaw and refreeze at the surface. For a sealed solid, the greater risk is moisture: warming and cooling drives air in and out past any imperfect seal, and water vapour travels with it and settles on a hygroscopic powder.
Three habits follow.
- Use a manual defrost unit for anything you intend to keep. A chest freezer also holds temperature better than an upright when it is opened, because cold air does not fall out of it.
- Never store reagents in the door, which is the warmest part of the cabinet and the part that cycles most.
- Defrost on a planned date, with the contents moved to another cold unit, rather than letting ice force an unplanned one.
The date of opening, not the date of purchase
An unopened vial has an intact closure and the headspace it was sealed with. Nothing has been admitted. The clock that matters to a stored reagent starts when that seal is first broken, because every entry afterwards admits room air, water vapour and whatever is on the transfer needle or the spatula.
This is why a purchase date on its own tells you little. A vial bought a year ago and never opened may be in better condition than one bought last month and entered twenty times. Record on the vial itself:
- the date of first opening
- the date of reconstitution, if the contents are now a solution
- the solvent used, and its own in use period as stated on its labelling
- the number of entries, where the protocol asks for it
A multiple entry solvent in a stoppered vial carries an in use period that begins at first puncture. A single use ampoule has none, because it is not intended to be entered twice. Honour whichever applies rather than the one that is more convenient.
What goes wrong
- A cabinet runs warm for a week and nobody knows, because nothing logs it.
- Reagents live in the freezer door, through the widest temperature swing in the building.
- A frost free unit cycles the stock several times a day for months.
- A vial carries a purchase date and no opening date, so its real history is unrecoverable.
- A solution is moved to a colder zone without being aliquoted first, guaranteeing more freeze thaw cycles.
Checklist
- Match the zone to the form: sealed solid, or solution.
- Manual defrost cabinets, contents in the main cabinet and not the door.
- A thermometer and a log for every cabinet that holds stock.
- Opening date and reconstitution date written on the vial.
- Aliquot before a solution goes into a freezer, never after.
Terms used here are collected in the glossary.