Water is the most used reagent in a laboratory and the least often specified. A protocol that says “add water” is ambiguous: the word covers at least six products that differ in what has been taken out of them and in whether anything in them is alive. This note sets out the common terms and where each belongs.

Purity and sterility answer different questions

Purity describes what is dissolved or suspended in the water: ions, organic carbon, particles and dissolved gases. Sterility describes whether viable microorganisms are present. The two are independent. Tap water run through an autoclave is sterile and still carries calcium, chloride and dissolved organic matter. Ultrapure water from a polishing unit is chemically clean and not sterile.

A third question sits beside both: endotoxin. Endotoxins are fragments of bacterial cell wall left behind after the organisms are dead. They survive autoclaving, so a sterile liquid can still contain them, and they are controlled only by a process specified for that purpose.

Distilled water

Distillation boils water and condenses the vapour, leaving behind salts, most metal ions, particles and anything else that does not travel with the steam. It does not remove substances that boil near water or carry over with it, including ammonia, carbon dioxide and some low molecular weight organics.

Deionised water

Deionisation passes water through ion exchange resins that swap dissolved cations for hydrogen ions and dissolved anions for hydroxide ions, which then combine as water. It removes ionic species efficiently and leaves behind everything uncharged: non-ionic organics, particles and microorganisms. A resin bed is also a damp surface with a very large area, so a neglected deioniser becomes a source of bacteria.

Purified water

Purified water names a grade rather than a process. In pharmacopoeial use it carries limits on conductivity and on organic carbon, and may be produced by distillation, ion exchange, reverse osmosis or a combination. Elsewhere the phrase tells you only that something was removed.

Ultrapure water

Ultrapure water, often called type 1, is purified water polished immediately before use, usually with a final ion exchange stage and a filter. Resistivity is the usual indicator, with 18.2 megohm cm at 25 degrees Celsius as the practical ceiling, alongside a low total organic carbon figure. ISO 3696 sets out grades of analytical laboratory water in comparable terms.

Its defining property is that it does not keep. With no buffering capacity, carbon dioxide from the air acidifies it within minutes, and it leaches ions from whatever container it stands in. It is drawn at the point of use, not stored on the bench.

Water for injection

Water for injection is a pharmacopoeial grade: purified water that additionally meets a limit for bacterial endotoxins. It is not automatically sterile, since that depends on how it was processed and packaged afterwards. The grade answers the endotoxin question, which none of the grades above address.

The name is the pharmacopoeial title of the grade and nothing more. It appears in this list so that the endotoxin limit can be compared with the other grades, in the same way conductivity is compared across the grades above.

Sterile water

Sterile water has been rendered free of viable microorganisms, normally by autoclaving or by filtration through a 0.22 micrometre filter followed by aseptic filling into a sealed container. The label says nothing about ionic purity. Once the closure is pierced the contents are no longer sterile, which is why sterile water is supplied in small sealed volumes.

Bacteriostatic water

Bacteriostatic water is sterile water containing a bacteriostatic agent, most commonly benzyl alcohol at 0.9 percent by weight. The agent does not sterilise anything. It inhibits the growth of organisms introduced when the closure is pierced, which is what allows a sealed vial to be entered more than once.

The preservative is therefore present in every solution made with it and can interfere with an analysis: benzyl alcohol carries an aromatic ring and absorbs in the ultraviolet.

Comparison

Grade Removes Does not address Typical bench use
Distilled salts, most ions, particles, non-volatiles volatiles, dissolved gases, microorganisms general rinsing, bulk solutions
Deionised ionic species organics, particles, microorganisms glassware rinsing, routine buffers
Purified ions and organic carbon to a defined limit microorganisms, endotoxin reagent preparation to a specification
Ultrapure (type 1) ions, organic carbon, particles microorganisms, endotoxin chromatography, trace analysis
Water for injection ions, organic carbon, endotoxin sterility, unless processed for it endotoxin sensitive work
Sterile viable microorganisms ions, organic carbon, endotoxin reconstituting reagents in sealed vials
Bacteriostatic viable microorganisms, then inhibits growth ions, organic carbon, endotoxin sealed vials entered more than once

What goes wrong

  • Treating “deionised” as a synonym for clean. A deioniser removes charge, not life. Water from a long idle line can carry a heavy microbial load while reading a perfectly good conductivity.
  • Storing ultrapure water. A bottle filled on Monday is ordinary purified water by Tuesday, with a lower pH and ions leached from the wall.
  • Ignoring the preservative. If an assay is sensitive to ultraviolet absorbing species or to alcohols, the solvent belongs in the method record.
  • Mixing grades within one series. A difference between two runs that is really a difference in solvent is almost impossible to diagnose afterwards.

A short checklist

  • Decide which of the three questions applies: purity, sterility, endotoxin.
  • Specify the grade in the protocol, not just the word water.
  • Draw ultrapure water fresh and use it immediately.
  • Keep one grade for one series, and record it on the label.

Where a solvent is bought in rather than produced on site, the grade is a statement on the label and on the accompanying documentation, not something to infer from the name on the bottle. Terms used here are defined in the glossary.